Main Menu
- Home
- Product Finder
- Calibration Systems
- Calibration Services
- Digital Sensing
- Industrial Vibration Calibration
- Modal and Vibration Testing
- Non-Destructive Testing
- Sound & Vibration Rental Program
- Learn
- About Us
- Contact Us
Similar in operation to Portable Vibration Calibrator, Model 9110D, Model C9110D, is a Portable Vibration Calibrator with charge amplifier input. The C9110D supports the calibration of both single-ended and differential charge accelerometers and is supplied with a 7/16-27 2-pin MIL to BNC cable for the calibration of high temperature turbine vibration sensors that have a differential output. Model C9110D is also ideal for calibrating ICP accelerometers, velocity transducers, proximity probes, and piezoresistive sensors over a wide operating frequency and amplitude range.
The unit is a compact, battery-powered, and completely self-contained vibration reference source, which can be conveniently used to calibrate individual sensors, vibration switches, and data collectors, as well as to validate the entire measurement channel of a condition monitoring or recording system. An integral precision quartz reference accelerometer and closed-loop level control gives the C9110D enhanced stability and superior vibration calibration over an extended 5 Hz to 10 kHz frequency range. The device is packaged in a rugged Pelican® Storm case.
CALROUTE firmware allows technicians to program repetitive calibration test points and pass/fail tolerances for both frequency response and linearity. The C9110D creates professional, ISO-17025 compliant calibration certificates that are fully customizable. The C9110D is supplied with an accuracy calibration certificate. Calibration of the C9110D is accredited to the ISO 17025 standard by the American Association of Laboratory Accreditation (A2LA).
View Datasheet | View Specification Sheet
US | Metric | |
---|---|---|
Frequency Range (operating) [1] | 5 Hz to 10 kHz | 300 to 600k CPM |
Maximum Amplitude (50 Hz, 10-gram payload) | 20 g pk | 196 m/s² pk |
Accuracy of Displayed Acceleration & Sensitivity Readout (10 Hz to 10 kHz) [2] [3] | ±3 % | |
Operating Battery Life [4] | 18 hours | |
Weight | 18 lb | 8.2 kg |
Sensor Mounting Platform Thread Size | ¼-28 |
[1] 100-gram payload
[2] Measured with 10-gram quartz reference accelerometer
[3] Calculated by measuring the % difference between the known sensitivity of a reference accelerometer as calibrated by laser primary system per
ISO 16063-11 and the measured sensitivity of same reference accelerometer when tested at the same points
[4] As shipped from factory in new condition and measured at 100 Hz, 1 g pk with 100-gram payload
The C9110D creates ISO 17025 compliant calibration reports via ISO 16063-21 methodology for the following vibration sensor types:
Each Portable Vibration Calibrator with Charge Amplifier Input Model C9110D is supplied with two mounting plates designed for the most popular 3 and 4-hole bolt patterns used on high temperature differential piezoelectric charge accelerometer designs and moving coil sensors. The following sensor models and series are supported: